Our high speed cell sorter (FACSAria) can perform 2-way and 4-way cell sorting.

Optic Specifications
Our FACSAria has 3 lasers: blue-488nm, red-633nm, and violet-407nm
A yellow laser of 561nm will be installed soon.
Please refer to the following table to select your Fluorophores. If you have to use any other fluorophores outside the following spectra, additional filter has to be purchased.


FACSAria Filter configuration
Lasers Longpass
Dichroic Mirror (nm)

Bandpass Filter (nm) Intended Dye
Blue (488nm) 735 780/60 PE-Cy7
  655 695/40 PerCP-Cy5.5
  595 610/20 PE-Texas Red
  556 575/26 PE
  502 530/30 FITC
Red (633nm) 735 780/60 APC-Cy7, APC-H7
  blank 660/20 APC
Violet (405nm) 502 530/30 Alexa Fluor 430
  blank 450/40 Cascade Blue
Pacific blue
Hoechst
DAPI
Alexa Fluor 405
Yellow (561nm) 735 780/60 PE-Cy7
  630 670/60 PE-Cy5
  600 610/20 mCherry, DsRed,
CherryTomato, PE,
Texas Red, PI

 

  582/15 PE (yellow-green)


Cells for Sorting
• Cells for sorting must be filtered using Falcon 12 x 75 mm polystyrene test tubes with cell strainer
    Snap cap (Falcon #2235)
. Filter your samples inside the culture hood right before bring to sort.
• Minimum sample volume is 0.5 ml.
• Cells should be concentrated to at 10- 20 million cells/ml for the high speed sorters. Ex. Lymphocytes
    – concentrate can be higher (up to 50 million cells/ml) Ex. Sticky/Adherent cells – concentrate at
    maximum 20 million cells/ml
• Place all the cell tubes and collecting tubes on ice

Bring Controls
1) A negative control: unstained sample or a sample stained only with the secondary Ab.
2) Single color controls. If you have samples stained with more than one fluorochrome per test tube,
      bring in a control sample stained with each fluorochrome individually. This is for compensation
      purposes.
      Ex. of controls for a simple GFP experiment:
             • Negative control = cells not expressing GFP
      Ex. of controls for a dual FITC-PE Experiment
             • Unlabeled control
             • FITC only control
             • PE only control

Collecting Media
1) Use appropriate media to collect cells:  put about 1 ml of collecting media in the collection tubes.
2) Use appropriate size of collection tubes.
      Ex. If you expect to collect 3 million cells, use 5 ml or 15 ml tube
      Ex. If you expect 100,000 cells, use a 1.5 ml eppendorf tube
3) Add antibiotics in the collecting media if sorted cells are to be cultured following cell sorting. Wash
      the collected cells once after sorting. The FACSAria cell sorting system could not be 100% sterilized
      unless this equipment is put inside a sterile hood.


The current setup for the cell sorter (FACSAria), which has no containment system for aerosols, prevents us from running samples that contain radioactivity or potentially infectious agents to humans.